:TRIGger Commands 31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 881
:TRIGger:HFReject
(see page 1138)
Command Syntax
:TRIGger:HFReject <value>
<value> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:HFReject command turns the high frequency reject filter off
and on. The high frequency reject filter adds a 50 kHz low- pass filter in
the trigger path to remove high frequency components from the trigger
waveform. Use this filter to remove high- frequency noise, such as AM or
FM broadcast stations, from the trigger path.
Query Syntax
:TRIGger:HFReject?
The :TRIGger:HFReject? query returns the current high frequency reject
filter mode.
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger[:EDGE]:REJect" on page 904