:MEASure Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 449
:MEASure:STATistics:MCOunt
(see page 1138)
Command Syntax
:MEASure:STATistics:MCOunt <setting>
<setting> ::= {INFinite | <count>}
<count> ::= 2 to 2000 in NR1 format
The :MEASure:STATistics:MCOunt command specifies the maximum
number of values used when calculating measurement statistics.
Query Syntax
:MEASure:STATistics:MCOunt?
The :MEASure:STATistics:MCOunt? query returns the current measurement
statistics max count setting.
Return Format
<setting><NL>
<setting> ::= {INF | <count>}
<count> ::= 2 to 2000
See Also • "Introduction to :MEASure Commands" on page 406
• ":MEASure:RESults" on page 438
• ":MEASure:STATistics" on page 446
• ":MEASure:STATistics:DISPlay" on page 447
• ":MEASure:STATistics:RSDeviation" on page 451
• ":MEASure:STATistics:RESet" on page 450
• ":MEASure:STATistics:INCRement" on page 448

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals