482 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
21 :MEASure Power Commands
:MEASure:OPOWer
(see page 1138)
Command Syntax
:MEASure:OPOWer
The :MEASure:OPOWer command installs an output power measurement on
screen.
Before sending this command or query, you must specify the channels
probing the input voltage, input current, output voltage, and output
current (using the :POWer:SIGNals:SOURce:VOLTage<i> and
:POWer:SIGNals:SOURce:CURRent<i> commands) and you must perform the
automated signals setup (using the :POWer:SIGNals:AUTosetup EFFiciency
command).
Query Syntax
:MEASure:OPOWer?
The :MEASure:OPOWer query returns the measured output power.
Return Format
<return_value><NL>
<return_value> ::= the output power value in NR3 format
See Also ":POWer:SIGNals:SOURce:VOLTage<i>" on page 582
":POWer:SIGNals:SOURce:CURRent<i>" on page 581
":POWer:SIGNals:AUTosetup" on page 564
":POWer:EFFiciency:APPLy" on page 535