1032 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
33 :WGEN Commands
:WGEN:VOLTage:OFFSet
(see page 1138)
Command Syntax
:WGEN:VOLTage:OFFSet <offset>
<offset> ::= offset in volts in NR3 format
The :WGEN:VOLTage:OFFSet command specifies the waveform's offset
voltage or the DC level. Use the :WGEN:VOLTage command to specify the
amplitude.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example,
an amplitude of 5 V and an offset of 1 V is the same as a high- level
voltage of 4 V and a low-level voltage of - 1 V.
Query Syntax
:WGEN:VOLTage:OFFSet?
The :WGEN:VOLTage:OFFSet? query returns the currently specified
waveform offset voltage.
Return Format
<offset><NL>
<offset> ::= offset in volts in NR3 format
See Also "Introduction to :WGEN Commands" on page 998
":WGEN:FUNCtion" on page 1007
":WGEN:VOLTage" on page 1029
":WGEN:VOLTage:HIGH" on page 1030
":WGEN:VOLTage:LOW" on page 1031