:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 735
:SBUS<n>:M1553:TRIGger:PATTern:DATA
(see page 1138)
Command Syntax
:SBUS<n>:M1553:TRIGger:PATTern:DATA <string>
<string> ::= "nn...n" where n ::= {0 |1|X}
The :SBUS<n>:M1553:TRIGger:PATTern:DATA command sets the 11 bits to
trigger on if the trigger type has been set to RTA11 (RTA + 11 Bits) using
the :SBUS<n>:M1553:TRIGger:TYPE command.
Query Syntax
:SBUS<n>:M1553:TRIGger:PATTern:DATA?
The :SBUS<n>:M1553:TRIGger:PATTern:DATA? query returns the current
11- bit setting.
Return Format
<string><NL>
<string> ::= "nn...n" where n ::= {0 |1|X}
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:M1553:TRIGger:TYPE" on page 737
":SBUS<n>:M1553:TRIGger:RTA" on page 736