466 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:WINDow
(see page 1138)
Command Syntax
:MEASure:WINDow <type>
<type> ::= {MAIN | ZOOM | AUTO}
When the zoomed time base is displayed, the :MEASure:WINDow command
lets you specify the measurement window:
• MAIN — the measurement window is the upper, Main window.
• ZOOM — the measurement window is the lower, Zoom window.
• AUTO — the measurement is attempted in the lower, Zoom window; if it
cannot be made there, the upper, Main window is used.
Query Syntax
:MEASure:WINDow?
The :MEASure:WINDow? query returns the current measurement window
setting.
Return Format
<type><NL>
<type> ::= {MAIN | ZOOM | AUTO}
See Also • "Introduction to :MEASure Commands" on page 406
• ":MEASure:SOURce" on page 444