:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 671
:SBUS<n>:FLEXray:AUTosetup
(see page 1138)
Command Syntax
:SBUS<n>:FLEXray:AUTosetup
The :SBUS<n>:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
Sets the selected source channel's impedance to 50 Ohms.
Sets the selected source channel' s probe attenuation to 10:1.
Sets the trigger level (on the selected source channel) to - 300 mV.
Turns on trigger Noise Reject.
Turns on Serial Decode.
Sets the trigger to the specified serial bus (n of SBUS<n>).
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:FLEXray:TRIGger" on page 679
":SBUS<n>:FLEXray:BAUDrate" on page 672
":TRIGger[:EDGE]:LEVel" on page 903
":SBUS<n>:FLEXray:SOURce" on page 678