:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 711
:SBUS<n>:IIC:TRIGger:PATTern:ADDRess
(see page 1138)
Command Syntax
:SBUS<n>:IIC:TRIGger:PATTern:ADDRess <value>
<value> ::= integer or <string>
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F}
The :SBUS<n>:IIC:TRIGger:PATTern:ADDRess command sets the address for
IIC data.The address can range from 0x00 to 0x7F (7- bit) or 0x3FF
(10- bit) hexadecimal. Use the don't care address (- 1 or 0xFFFFFFFF) to
ignore the address value.
Query Syntax
:SBUS<n>:IIC:TRIGger:PATTern:ADDRess?
The :SBUS<n>:IIC:TRIGger:PATTern:ADDRess? query returns the current
address for IIC data.
Return Format
<value><NL>
<value> ::= integer
See Also • "Introduction to :TRIGger Commands" on page 877
• ":SBUS<n>:IIC:TRIGger:PATTern:DATA" on page 712
• ":SBUS<n>:IIC:TRIGger:PATTern:DATa2" on page 713
• ":SBUS<n>:IIC:TRIGger[:TYPE]" on page 715

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals