460 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:VMIN
(see page 1138)
Command Syntax
:MEASure:VMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VMIN command installs a screen measurement and starts a
minimum vertical value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VMIN? [<source>]
The :MEASure:VMIN? query measures and outputs the minimum vertical
value present on the selected waveform.
Return Format
<value><NL>
<value> ::= minimum vertical value of the selected waveform in
NR3 format
See Also "Introduction to :MEASure Commands" on page 406
":MEASure:SOURce" on page 444
":MEASure:VBASe" on page 458
":MEASure:VMAX" on page 459
":MEASure:VPP" on page 461