:SYSTem Commands 29
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 855
:SYSTem:ERRor
(see page 1138)
Query Syntax
:SYSTem:ERRor?
The :SYSTem:ERRor? query outputs the next error number and text from
the error queue. The instrument has an error queue that is 30 errors deep
and operates on a first- in, first- out basis. Repeatedly sending the
:SYSTem:ERRor? query returns the errors in the order that they occurred
until the queue is empty. Any further queries then return zero until
another error occurs.
Return Format
<error number>,<error string><NL>
<error number> ::= an integer error code in NR1 format
<error string> ::= quoted ASCII string containing the error message
Error messages are listed in Chapter 36, “Error Messages,” starting on
page 1097.
See Also "Introduction to :SYSTem Commands" on page 852
"*ESR (Standard Event Status Register)" on page 168
"*CLS (Clear Status)" on page 165

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals