:TIMebase Commands 30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 873
:TIMebase:WINDow:POSition
(see page 1138)
Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the zoomed (delayed)
view reference point in NR3 format
The :TIMebase:WINDow:POSition command sets the horizontal position in
the zoomed (delayed) view of the main sweep. The main sweep range and
the main sweep horizontal position determine the range for this command.
The value for this command must keep the zoomed view window within
the main sweep range.
Query Syntax :TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal
window position setting in the zoomed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also "Introduction to :TIMebase Commands" on page 866
":TIMebase:MODE" on page 867
":TIMebase:POSition" on page 868
":TIMebase:RANGe" on page 869
":TIMebase:SCALe" on page 871
":TIMebase:WINDow:RANGe" on page 874
":TIMebase:WINDow:SCALe" on page 875