:MEASure Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 409
:MEASure:AREa
(see page 1138)
Command Syntax
:MEASure:AREa [<interval>][,][<source>]
<interval> ::= {CYCLe | DISPlay}
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:AREa command installs an area measurement on screen.
Area measurements show the area between the waveform and the ground
level.
The <interval> option lets you specify the measurement interval: either an
integral number of cycles, or the full screen. If <interval> is not specified,
DISPlay is implied.
Query Syntax
:MEASure:AREa? [<interval>][,][<source>]
The :MEASure:AREa? query measures and returns the area value.
Return Format
<value><NL>
<value> ::= the area value in volt-seconds in NR3 format
See Also • "Introduction to :MEASure Commands" on page 406
• ":MEASure:SOURce" on page 444
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals