842 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :SEARch Commands
:SEARch:SERial:M1553:RTA
(see page 1138)
Command Syntax
:SEARch:SERial:M1553:RTA <value>
<value> ::= 5-bit integer in decimal, <hexadecimal>, <binary>,
or <string> from 0-31
<hexadecimal> ::= #Hnn where n ::= {0,..,9|A,..,F}
<binary> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n::= {0,..,9|A,..,F}
The :SEARch:SERial:M1553:RTA command specifies the Remote Terminal
Address (RTA) value when searching for MIL- STD- 1553 events.
Query Syntax
:SEARch:SERial:M1553:RTA?
The :SEARch:SERial:M1553:RTA? query returns the current Remote
Terminal Address value setting.
Return Format
<value><NL>
<value> ::= 5-bit integer in decimal from 0-31
See Also Chapter 28, “:SEARch Commands,” starting on page 777