974 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
32 :WAVeform Commands
:WAVeform:POINts:MODE
(see page 1138)
Command Syntax
:WAVeform:POINts:MODE <points_mode>
<points_mode> ::= {NORMal | MAXimum | RAW}
The :WAVeform:POINts:MODE command sets the data record to be
transferred with the :WAVeform:DATA? query.
For the analog or digital sources, there are two different records that can
be transferred:
• The first is the raw acquisition record. The maximum number of points
available in this record is returned by the :ACQuire:POINts? query. The
raw acquisition record can only be transferred when the oscilloscope is
not running and can only be retrieved from the analog or digital
sources.
• The second is referred to as the measurement record and is a
62,500- point (maximum) representation of the raw acquisition record.
The measurement record can be retrieved from any source.
If the <points_mode> is NORMal the measurement record is retrieved.
If the <points_mode> is RAW, the raw acquisition record is used. Under
some conditions, such as when the oscilloscope is running, this data
record is unavailable.
If the <points_mode> is MAXimum, whichever record contains the
maximum amount of points is used. Usually, this is the raw acquisition
record. But, if the raw acquisition record is unavailable (for example,
when the oscilloscope is running), the measurement record may have more
data. If data is being retrieved as the oscilloscope is stopped and as the
data displayed is changing, the data being retrieved can switch between
the measurement and raw acquisition records.
Considerations
for MAXimum or
RAW data
retrieval
• The instrument must be stopped (see the :STOP command (see
page 223) or the :DIGitize command (see page 199) in the root
subsystem) in order to return more than the measurement record.
• :TIMebase:MODE must be set to MAIN.
• :ACQuire:TYPE must be set to NORMal or HRESolution.
• MAXimum or RAW will allow up to 4,000,000 points to be returned. The
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current
settings.
Query Syntax
:WAVeform:POINts:MODE?