:WAVeform Commands 32
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 989
:WAVeform:XINCrement
(see page 1138)
Query Syntax
:WAVeform:XINCrement?
The :WAVeform:XINCrement? query returns the x- increment value for the
currently specified source. This value is the time difference between
consecutive data points in seconds.
Return Format
<value><NL>
<value> ::= x-increment in the current preamble in 64-bit
floating point NR3 format
See Also "Introduction to :WAVeform Commands" on page 961
":WAVeform:PREamble" on page 976
Example Code "Example Code" on page 977