758 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:UART:BAUDrate
(see page 1138)
Command Syntax
:SBUS<n>:UART:BAUDrate <baudrate>
<baudrate> ::= integer from 100 to 8000000
The :SBUS<n>:UART:BAUDrate command selects the bit rate (in bps) for
the serial decoder and/or trigger when in UART mode. The baud rate can
be set from 100 b/s to 8 Mb/s.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:SBUS<n>:UART:BAUDrate?
The :SBUS<n>:UART:BAUDrate? query returns the current UART baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 100 to 8000000
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:MODE" on page 886
":SBUS<n>:UART:TRIGger:TYPE" on page 774