1004 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
33 :WGEN Commands
:WGEN:ARBitrary:INTerpolate
(see page 1138)
Command Syntax
:WGEN:ARBitrary:INTerpolate {{0 | OFF} | {1 | ON}}
The :WGEN:ARBitrary:INTerpolate command enables or disables the
Interpolation control.
Interpolation specifies how lines are drawn between arbitrary waveform
points:
When ON, lines are drawn between points in the arbitrary waveform.
Voltage levels change linearly between one point and the next.
When OFF, all line segments in the arbitrary waveform are horizontal.
The voltage level of one point remains until the next point.
Query Syntax
:WGEN:ARBitrary:INTerpolate?
The :WGEN:ARBitrary:INTerpolate query returns the current interpolation
setting.
Return Format
{0 | 1}
See Also ":WGEN:ARBitrary:DATA" on page 1000
":WGEN:ARBitrary:DATA:DAC" on page 1003
":SAVE:ARBitrary[:STARt]" on page 608
":RECall:ARBitrary[:STARt]" on page 599