EasyManua.ls Logo

Agilent Technologies InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
476 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
21 :MEASure Power Commands
:MEASure:EFFiciency
(see page 1138)
Command Syntax
:MEASure:EFFiciency
The :MEASure:EFFiciency command installs an efficiency (output power /
input power) measurement on screen.
Before sending this command or query, you must specify the channels
probing the input voltage, input current, output voltage, and output
current (using the :POWer:SIGNals:SOURce:VOLTage<i> and
:POWer:SIGNals:SOURce:CURRent<i> commands) and you must perform the
automated signals setup (using the :POWer:SIGNals:AUTosetup EFFiciency
command).
Query Syntax
:MEASure:EFFiciency?
The :MEASure:EFFiciency query returns the measured efficiency as a
percent value.
Return Format
<return_value><NL>
<return_value> ::= percent value in NR3 format
See Also ":POWer:SIGNals:SOURce:VOLTage<i>" on page 582
":POWer:SIGNals:SOURce:CURRent<i>" on page 581
":POWer:SIGNals:AUTosetup" on page 564
":POWer:EFFiciency:APPLy" on page 535

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Related product manuals