EasyManuals Logo
Home>Agilent Technologies>Test Equipment>InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series Programmer's Guide

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #476 background imageLoading...
Page #476 background image
476 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
21 :MEASure Power Commands
:MEASure:EFFiciency
(see page 1138)
Command Syntax
:MEASure:EFFiciency
The :MEASure:EFFiciency command installs an efficiency (output power /
input power) measurement on screen.
Before sending this command or query, you must specify the channels
probing the input voltage, input current, output voltage, and output
current (using the :POWer:SIGNals:SOURce:VOLTage<i> and
:POWer:SIGNals:SOURce:CURRent<i> commands) and you must perform the
automated signals setup (using the :POWer:SIGNals:AUTosetup EFFiciency
command).
Query Syntax
:MEASure:EFFiciency?
The :MEASure:EFFiciency query returns the measured efficiency as a
percent value.
Return Format
<return_value><NL>
<return_value> ::= percent value in NR3 format
See Also • ":POWer:SIGNals:SOURce:VOLTage<i>" on page 582
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 581
• ":POWer:SIGNals:AUTosetup" on page 564
• ":POWer:EFFiciency:APPLy" on page 535

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals