928 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31 :TRIGger Commands
:TRIGger:RUNT:POLarity
(see page 1138)
Command Syntax
:TRIGger:RUNT:POLarity <polarity>
<polarity> ::= {POSitive | NEGative | EITHer}
The :TRIGger:RUNT:POLarity command sets the polarity for the runt
trigger:
POSitive — positive runt pulses.
NEGative — negative runt pulses.
EITHer — either positive or negative runt pulses.
Query Syntax
:TRIGger:RUNT:POLarity?
The :TRIGger:RUNT:POLarity? query returns the runt trigger polarity.
Return Format
<polarity><NL>
<polarity> ::= {POS | NEG | EITH}
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:MODE" on page 886
":TRIGger:LEVel:HIGH" on page 884
":TRIGger:LEVel:LOW" on page 885
":TRIGger:RUNT:SOURce" on page 930