:TIMebase Commands 30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 871
:TIMebase:SCALe
(see page 1138)
Command Syntax
:TIMebase:SCALe <scale_value>
<scale_value> ::= time/div in seconds in NR3 format
The :TIMebase:SCALe command sets the horizontal scale or units per
division for the main window.
Query Syntax
:TIMebase:SCALe?
The :TIMebase:SCALe? query returns the current horizontal scale setting in
seconds per division for the main window.
Return Format
<scale_value><NL>
<scale_value> ::= time/div in seconds in NR3 format
See Also • "Introduction to :TIMebase Commands" on page 866
• ":TIMebase:RANGe" on page 869
• ":TIMebase:WINDow:SCALe" on page 875
• ":TIMebase:WINDow:RANGe" on page 874