:TRIGger Commands 31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 929
:TRIGger:RUNT:QUALifier
(see page 1138)
Command Syntax
:TRIGger:RUNT:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | NONE}
The :TRIGger:RUNT:QUALifier command selects the qualifier used for
specifying runt pulse widths:
• GREaterthan — triggers on runt pulses whose width is greater than the
:TRIGger:RUNT:TIME.
• LESSthan — triggers on runt pulses whose width is less than the
:TRIGger:RUNT:TIME.
• NONE — triggers on runt pulses of any width.
Query Syntax
:TRIGger:RUNT:QUALifier?
The :TRIGger:RUNT:QUALifier? query returns the runt trigger qualifier
setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS NONE}
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:MODE" on page 886
• ":TRIGger:RUNT:TIME" on page 931