726 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:LIN:TRIGger:ID
(see page 1138)
Command Syntax
:SBUS<n>:LIN:TRIGger:ID <value>
<value> ::= 7-bit integer in decimal, <nondecimal>, or <string>
from 0-63 or 0x00-0x3f
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :SBUS<n>:LIN:TRIGger:ID command defines the LIN identifier searched
for in each CAN message when the LIN trigger mode is set to frame ID.
Setting the ID to a value of "- 1" results in "0xXX" which is equivalent to
all IDs.
Query Syntax
:SBUS<n>:LIN:TRIGger:ID?
The :SBUS<n>:LIN:TRIGger:ID? query returns the current LIN identifier
setting.
Return Format
<value><NL>
<value> ::= integer in decimal
Errors • "- 241, Hardware missing" on page 1099
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:MODE" on page 886
• ":SBUS<n>:LIN:TRIGger" on page 725
• ":SBUS<n>:LIN:SOURce" on page 722

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals