488 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
21 :MEASure Power Commands
:MEASure:TRESponse
(see page 1138)
Command Syntax
:MEASure:TRESponse [<source>]
<source> ::= {CHANnel<n>| FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:TRESponse command installs a transient response time
measurement on screen.
The <source> parameter is the channel probing the output voltage. This
source can also be specified by the :MEASure:SOURce command.
Transient response time = t2 – t1, where:
t1 = The first time a voltage waveform exits the settling band.
t2 = The last time it enters into the settling band.
Settling band = +/- overshoot % of the steady state output voltage.
Query Syntax
:MEASure:TRESponse? [<source>]
The :MEASure:TRESponse query returns the measured transient response
time.
Return Format
<return_value><NL>
<return_value> ::= time in seconds for the overshoot to settle back
into the band in NR3 format
See Also ":MEASure:SOURce" on page 444
":POWer:TRANsient:APPLy" on page 591