:POWer Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 557
:POWer:PSRR:APPLy
(see page 1138)
Command Syntax
:POWer:PSRR:APPLy
The :POWer:PSRR:APPLy command applies the power supply rejection ratio
(PSRR) analysis.
The Power Supply Rejection Ratio (PSRR) test is used to determine how
well a voltage regulator rejects ripple noise over different frequency range.
This analysis provides a signal from the oscilloscope's waveform generator
that sweeps its frequency. This signal is used to inject ripple to the DC
voltage that feeds the voltage regulator.
The AC RMS ratio of the input over the output is measured and is plotted
over the range of frequencies.
See Also • ":POWer:PSRR:FREQuency:MAXimum" on page 558
• ":POWer:PSRR:FREQuency:MINimum" on page 559
• ":POWer:PSRR:RMAXimum" on page 560

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals