EasyManua.ls Logo

Agilent Technologies InfiniiVision 3000 X-Series - Power:psrr:apply

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:POWer Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 557
:POWer:PSRR:APPLy
(see page 1138)
Command Syntax
:POWer:PSRR:APPLy
The :POWer:PSRR:APPLy command applies the power supply rejection ratio
(PSRR) analysis.
The Power Supply Rejection Ratio (PSRR) test is used to determine how
well a voltage regulator rejects ripple noise over different frequency range.
This analysis provides a signal from the oscilloscope's waveform generator
that sweeps its frequency. This signal is used to inject ripple to the DC
voltage that feeds the voltage regulator.
The AC RMS ratio of the input over the output is measured and is plotted
over the range of frequencies.
See Also ":POWer:PSRR:FREQuency:MAXimum" on page 558
":POWer:PSRR:FREQuency:MINimum" on page 559
":POWer:PSRR:RMAXimum" on page 560

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Related product manuals