:WAVeform Commands 32
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 967
:WAVeform:BYTeorder
(see page 1138)
Command Syntax
:WAVeform:BYTeorder <value>
<value> ::= {LSBFirst | MSBFirst}
The :WAVeform:BYTeorder command sets the output sequence of the WORD
data.
• MSBFirst — sets the most significant byte to be transmitted first.
• LSBFirst — sets the least significant byte to be transmitted first.
This command affects the transmitting sequence only when
:WAVeform:FORMat WORD is selected.
The default setting is MSBFirst.
Query Syntax
:WAVeform:BYTeorder?
The :WAVeform:BYTeorder query returns the current output sequence.
Return Format
<value><NL>
<value> ::= {LSBF | MSBF}
See Also • "Introduction to :WAVeform Commands" on page 961
• ":WAVeform:DATA" on page 969
• ":WAVeform:FORMat" on page 971
• ":WAVeform:PREamble" on page 976
Example Code • "Example Code" on page 982
• "Example Code" on page 977

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals