:CHANnel<n> Commands 10
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 275
:CHANnel<n>:PROBe:STYPe
(see page 1138)
Command Syntax
:CHANnel<n>:PROBe:STYPe <signal type>
<signal type> ::= {DIFFerential | SINGle}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:PROBe:STYPe command sets the channel probe signal
type (STYPe) to differential or single- ended when using the 113xA Series
probes and determines how offset is applied.
When single- ended is selected, the :CHANnel<n>:OFFset command changes
the offset value of the probe amplifier. When differential is selected, the
:CHANnel<n>:OFFset command changes the offset value of the channel
amplifier.
Query Syntax
:CHANnel<n>:PROBe:STYPe?
The :CHANnel<n>:PROBe:STYPe? query returns the current probe signal
type setting for the selected channel.
Return Format
<signal type><NL>
<signal type> ::= {DIFF | SING}
See Also • "Introduction to :CHANnel<n> Commands" on page 262
• ":CHANnel<n>:OFFSet" on page 270
NOTE
This command is valid only for the 113xA Series probes.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals