:WGEN Commands 33
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 1013
:WGEN:MODulation:AM:DEPTh
(see page 1138)
Command Syntax
:WGEN:MODulation:AM:DEPTh <percent>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used
by the modulation. For example, a depth setting of 80% causes the output
amplitude to vary from 10% to 90% (90% – 10% = 80%) of the original
amplitude as the modulating signal goes from its minimum to maximum
amplitude.
Query Syntax
:WGEN:MODulation:AM:DEPTh?
The :WGEN:MODulation:AM:DEPTh? query returns the AM depth
percentage setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also • ":WGEN:MODulation:AM:FREQuency" on page 1014
• ":WGEN:MODulation:FM:DEViation" on page 1015
• ":WGEN:MODulation:FM:FREQuency" on page 1016
• ":WGEN:MODulation:FSKey:FREQuency" on page 1017
• ":WGEN:MODulation:FSKey:RATE" on page 1018
• ":WGEN:MODulation:FUNCtion" on page 1019
• ":WGEN:MODulation:FUNCtion:RAMP:SYMMetry" on page 1020
• ":WGEN:MODulation:STATe" on page 1022
• ":WGEN:MODulation:TYPE" on page 1023