180 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
5 Common (*) Commands
*SRE (Service Request Enable)
(see page 1138)
Command Syntax
*SRE <mask>
<mask> ::= integer with values defined in the following table.
The *SRE command sets the bits in the Service Request Enable Register.
The Service Request Enable Register contains a mask value for the bits to
be enabled in the Status Byte Register. A one in the Service Request
Enable Register enables the corresponding bit in the Status Byte Register.
A zero disables the bit.
)URP
2SHUDWLRQ
6WDWXV
5HJLVWHUV

23(5 75*0$9(6%
546
066
75*
5HJ
7(5" 7ULJJHU(YHQW5HJLVWHU
65(
65("
67%" 6WDWXV%\WH5HJLVWHU
6HUYLFH5HTXHVW(QDEO
0DVN5HJLVWHU
2XWSXW
4XHXH
654
)URP
6WDQGDUG
(YHQW
6WDWXV
5HJLVWHUV
865
06*
6HUYLFH5HTXHVW
25

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals