*TST (Self Test)
(see page 1138)
Query Syntax
*TST?
The *TST? query performs a self- test on the instrument. The result of the
test is placed in the output queue. A zero indicates the test passed and a
non- zero indicates the test failed. If the test fails, refer to the
troubleshooting section of the Service Guide.
Return Format
<result><NL>
<result> ::= 0 or non-zero value; an integer in NR1 format
See Also • "Introduction to Common (*) Commands" on page 163