230 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
7 :ACQuire Commands
:ACQuire:COUNt
(see page 1138)
Command Syntax
:ACQuire:COUNt <count>
<count> ::= integer in NR1 format
In averaging mode, the :ACQuire:COUNt command specifies the number of
values to be averaged for each time bucket before the acquisition is
considered to be complete for that time bucket. When :ACQuire:TYPE is set
to AVERage, the count can be set to any value from 2 to 65536.
Query Syntax
:ACQuire:COUNT?
The :ACQuire:COUNT? query returns the currently selected count value for
averaging mode.
Return Format
<count_argument><NL>
<count_argument> ::= an integer from 2 to 65536 in NR1 format
See Also "Introduction to :ACQuire Commands" on page 227
":ACQuire:TYPE" on page 239
":DIGitize" on page 199
":WAVeform:COUNt" on page 968
NOTE
The :ACQuire:COUNt 1 command has been deprecated. The AVERage acquisition type with
a count of 1 is functionally equivalent to the HRESolution acquisition type; however, you
should select the high-resolution acquisition mode with the :ACQuire:TYPE HRESolution
command instead.