306 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
13 :DISPlay Commands
:DISPlay:LABList
(see page 1138)
Command Syntax
:DISPlay:LABList <binary block data>
<binary block> ::= an ordered list of up to 75 labels, a maximum of 10
characters each, separated by newline characters.
The :DISPlay:LABList command adds labels to the label list. Labels are
added in alphabetical order.
Query Syntax
:DISPlay:LABList?
The :DISPlay:LABList? query returns the label list.
Return Format
<binary block><NL>
<binary block> ::= an ordered list of up to 75 labels, a maximum of 10
characters each, separated by newline characters.
See Also • "Introduction to :DISPlay Commands" on page 298
• ":DISPlay:LABel" on page 305
• ":CHANnel<n>:LABel" on page 269
• ":DIGital<d>:LABel" on page 292
• ":BUS<n>:LABel" on page 248
NOTE
Labels that begin with the same alphabetic base string followed by decimal digits are
considered duplicate labels. Duplicate labels are not added to the label list. For example, if
label "A0" is in the list and you try to add a new label called "A123456789", the new label is
not added.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals