Command reference
R&S
®
ZNB/ZNBT
1303User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]SEGMent<Seg>:DEFine <StartFreq>, <StopFreq>, <Points>,
<Power>, <SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[,
<LO>, <Selectivity>, <FreqSweepMode>]
Creates or re-defines a sweep segment no. <Seg> with specific channel settings.
Entry of the first seven numeric parameters is mandatory; no default values are provi-
ded. All settings except <LO> can be changed for existing segments using other com-
mands of the [SENSe<Ch>:]SEGMent<Seg>... subsystem.
Note: Use [SENSe<Ch>:]SEGMent<Seg>:ADD to create a segment with default
channel settings. Use [SENSe<Ch>:]SEGMent<Seg>:INSert (no query) to insert a
new segment into the current segment list.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If segment number <Seg> already exists, it is replaced by the
new segment.
Parameters:
<StartFreq> Start frequency of the segment; see [SENSe<Ch>:
]SEGMent<Seg>:FREQuency:STARt.
Default unit: Hz
<StopFreq> Stop frequency of the segment; see [SENSe<Ch>:
]SEGMent<Seg>:FREQuency:STOP.
Default unit: Hz
<Points> Number of sweep points in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:SWEep:POINts.
<Power> Internal source power in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:POWer[:LEVel].
Default unit: dBm
<SegmentTime> Duration of the sweep in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:SWEep:TIME. In the setting [SENSe<Ch>:
]SEGMent<Seg>:INSert:SELect DWELl, this parameter is
replaced by <MeasDelay>.
Range: Depending on other channel settings. AUTO acti-
vates automatic sweep time setting in the segment,
which is equivalent to the minimum sweep time
possible.
Default unit: s
<MeasDelay> Delay for each partial measurement in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime,
this parameter is replaced by <SegmentTime>.
SCPI command reference