134 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
4 Commands Quick Reference
Table 39 :SEARch:SERial:CAN Commands Summary
Command Query Options and Query Returns
:SEARch:SERial:CAN:MO
DE <value> (see
page 809)
:SEARch:SERial:CAN:MO
DE? (see page 809)
<value> ::= {DATA | IDData |
IDEither | IDRemote | ALLerrors |
OVERload | ERRor}
:SEARch:SERial:CAN:PA
TTern:DATA <string>
(see page 810)
:SEARch:SERial:CAN:PA
TTern:DATA? (see
page 810)
<string> ::= "0xnn...n" where n
::= {0,..,9 | A,..,F | X} for
hexadecimal
:SEARch:SERial:CAN:PA
TTern:DATA:LENGth
<length> (see
page 811)
:SEARch:SERial:CAN:PA
TTern:DATA:LENGth?
(see page 811)
<length> ::= integer from 1 to 8
in NR1 format
:SEARch:SERial:CAN:PA
TTern:ID <string>
(see page 812)
:SEARch:SERial:CAN:PA
TTern:ID? (see
page 812)
<string> ::= "0xnn...n" where n
::= {0,..,9 | A,..,F | X} for
hexadecimal
:SEARch:SERial:CAN:PA
TTern:ID:MODE <value>
(see page 813)
:SEARch:SERial:CAN:PA
TTern:ID:MODE? (see
page 813)
<value> ::= {STANdard | EXTended}
Table 40 :SEARch:SERial:FLEXray Commands Summary
Command Query Options and Query Returns
:SEARch:SERial:FLEXra
y:CYCLe <cycle> (see
page 815)
:SEARch:SERial:FLEXra
y:CYCLe? (see
page 815)
<cycle> ::= {ALL | <cycle #>}
<cycle #> ::= integer from 0-63
:SEARch:SERial:FLEXra
y:DATA <string> (see
page 816)
:SEARch:SERial:FLEXra
y:DATA? (see
page 816)
<string> ::= "0xnn...n" where n
::= {0,..,9 | A,..,F | X }
:SEARch:SERial:FLEXra
y:DATA:LENGth
<length> (see
page 817)
:SEARch:SERial:FLEXra
y:DATA:LENGth? (see
page 817)
<length> ::= integer from 1 to 12
in NR1 format
:SEARch:SERial:FLEXra
y:FRAMe <frame id>
(see page 818)
:SEARch:SERial:FLEXra
y:FRAMe? (see
page 818)
<frame_id> ::= {ALL | <frame #>}
<frame #> ::= integer from 1-2047
:SEARch:SERial:FLEXra
y:MODE <value> (see
page 819)
:SEARch:SERial:FLEXra
y:MODE? (see
page 819)
<value> := {FRAMe | CYCLe | DATA
| HERRor | FERRor | AERRor}