166 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
5 Common (*) Commands
*ESE (Standard Event Status Enable)
(see page 1138)
Command Syntax
*ESE <mask_argument>
<mask_argument> ::= integer from 0 to 255
The *ESE common command sets the bits in the Standard Event Status
Enable Register. The Standard Event Status Enable Register contains a
mask value for the bits to be enabled in the Standard Event Status
Register. A "1" in the Standard Event Status Enable Register enables the
corresponding bit in the Standard Event Status Register. A zero disables
the bit.
Query Syntax
*ESE?
Table 65 Standard Event Status Enable (ESE)
Bit Name Description When Set (1 = High = True), Enables:
7 PON Power On Event when an OFF to ON transition occurs.
6 URQ User Request Event when a front-panel key is pressed.
5 CME Command Error Event when a command error is detected.
4 EXE Execution Error Event when an execution error is detected.
3 DDE Device Dependent Error Event when a device-dependent error is detected.
2 QYE Query Error Event when a query error is detected.
1 RQL Request Control Event when the device is requesting control. (Not
used.)
0 OPC Operation Complete Event when an operation is complete.
To ESB bit in
Statu s Byte
Register
Standard Event
Status Enable
(Mask) Register
Standard Event
Status Register
*ESR?
*ESE
*ESE?
PON URQ EXE DDE QYE RQL OPCCM E

OR
10234567
128 64 16 8 4 2 132