EasyManua.ls Logo

Agilent Technologies InfiniiVision 3000 X-Series - Page 224

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
224 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
6 Root (:) Commands
:TER (Trigger Event Register)
(see page 1138)
Query Syntax
:TER?
The :TER? query reads the Trigger Event Register. After the Trigger Event
Register is read, it is cleared. A one indicates a trigger has occurred. A
zero indicates a trigger has not occurred.
The Trigger Event Register is summarized in the TRG bit of the Status
Byte Register (STB). A Service Request (SRQ) can be generated when the
TRG bit of the Status Byte transitions, and the TRG bit is set in the
Service Request Enable register. The Trigger Event Register must be
cleared each time you want a new service request to be generated.
Return Format
<value><NL>
<value> ::= {1 | 0}; a 16-bit integer in NR1 format.
See Also "Introduction to Root (:) Commands" on page 190
"*SRE (Service Request Enable)" on page 180
"*STB (Read Status Byte)" on page 182

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Related product manuals