s1
241
Agilent InfiniiVision 3000 X-Series Oscilloscopes
Programmer's Guide
8
:BUS<n> Commands
Control all oscilloscope functions associated with buses made up of digital
channels. See "Introduction to :BUS<n> Commands" on page 242.
Table 80 :BUS<n> Commands Summary
Command Query Options and Query Returns
:BUS<n>:BIT<m> {{0 |
OFF} | {1 | ON}} (see
page 243)
:BUS<n>:BIT<m>? (see
page 243)
{0 | 1}
<n> ::= 1 or 2; an integer in NR1
format
<m> ::= 0-15; an integer in NR1
format
:BUS<n>:BITS
<channel_list>, {{0 |
OFF} | {1 | ON}} (see
page 244)
:BUS<n>:BITS? (see
page 244)
<channel_list>, {0 | 1}
<channel_list> ::= (@<m>,<m>:<m>
...) where "," is separator and
":" is range
<n> ::= 1 or 2; an integer in NR1
format
<m> ::= 0-15; an integer in NR1
format
:BUS<n>:CLEar (see
page 246)
n/a <n> ::= 1 or 2; an integer in NR1
format
:BUS<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
page 247)
:BUS<n>:DISPlay? (see
page 247)
{0 | 1}
<n> ::= 1 or 2; an integer in NR1
format

Table of Contents