:MEASure Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 415
Return Format for <meas_spec> = DELay:
{ <edge_spec1> | <edge_spec2> | <edge_spec1>,<edge_spec2>} <NL>
for <meas_spec> = THResholds and <threshold mode> = PERCent:
THR,PERC,<upper>,<middle>,<lower><NL>
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
and lower threshold percentage values
between Vbase and Vtop in NR3 format.
for <meas_spec> = THResholds and <threshold mode> = ABSolute:
THR,ABS,<upper>,<middle>,<lower><NL>
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
and lower threshold voltages in NR3
format.
for <threshold spec> = STANdard:
THR,PERC,+90.0,+50.0,+10.0
See Also "Introduction to :MEASure Commands" on page 406
":MEASure:DELay" on page 416
":MEASure:SOURce" on page 444
":CHANnel<n>:RANGe" on page 277
":CHANnel<n>:SCALe" on page 278
":CHANnel<n>:PROBe" on page 271
":CHANnel<n>:UNITs" on page 279