222 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
6 Root (:) Commands
:STATus
(see page 1138)
Query Syntax
:STATus? <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | SBUS{1 | 2} | WMEMory<r>}
for the DSO models
<source> ::= {CHANnel<n> | DIGital<d> | POD{1 | 2}
| BUS{1 | 2} | FUNCtion | MATH | SBUS{1 | 2} | WMEMory<r>}
for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :STATus? query reports whether the channel, function, or serial
decode bus specified by <source> is displayed.
Return Format
<value><NL>
<value> ::= {1 | 0}
See Also "Introduction to Root (:) Commands" on page 190
":BLANk" on page 198
":CHANnel<n>:DISPlay" on page 266
":DIGital<d>:DISPlay" on page 291
":FUNCtion:DISPlay" on page 333
":POD<n>:DISPlay" on page 525
":WMEMory<r>:DISPlay" on page 1036
":VIEW" on page 225
NOTE
MATH is an alias for FUNCtion.