EasyManua.ls Logo

ETC Cobalt 10 - Device Times - Introduction; Percent Times, or Seconds

ETC Cobalt 10
1076 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cobalt 8.1 - 20181029
641
Device Times - Introduction
A Moving Device can have Times and Delays in the following ways.
No time - follows the Sequence Step In-time.
Parameter group times - an attribute time for FOCUS, COLOR or BEAM parameters.
FCB times - Same as above, but there is a shortcut for setting all three
Device parameter times - A time for each attribute of a Moving Device.
NOTE
All times are set to the Preset in the A or B field depending on the setting of
the parameter "Set Times To Field" in the Cobalt Settings.
All attribute times are stored with a Preset.
There is a default Attribute Time that is used when Moving Devices are positioned manually or with
GoInB. See Moving Device Times - Default Attribute Time.
Percent times, or seconds?
Times for Moving Devices are set in seconds (absolute) or as a percentage of the IN time of a
Sequence Step.
The default setting is percent. This means that all FCB-Times will equal the IN time of a Sequence
Step, and all FCB-Delay Times will execute as a percentage of the IN time of a Sequence Step.
In the example above, the FCB Times are 100% of the In time (5 seconds).

Table of Contents

Related product manuals