-- 8 0 --
ACQuire:NUMACq?
Description
This query--only command returns the number of waveform acquisi-
tions that have occurred since starting acquisition with the AC-
Quire:STATE RUN command. This value is reset to zero when any
acquisition, horizontal, or vertical arguments that affect the waveform
are changed. The maximum number of acquisitions that can be counted
is
. The instrument stops counting when this number is reached.
This is the same value that displays in the upper center of the screen
when the acquisition system is stopped.
Group
Acquisition
Related Commands
ACQuire:STATE
Syntax
ACQuire:NUMACq?
Example
ACQuire:NUMACq?
This command might return
:ACQUIRE:NUMACQ 350
indicating that 350 acquisitions have occurred since executing an AC-
Quire:STATE RUN command.