TDS5000B Online Programmer Guide
-- 7 7 4 --
TRIGger:A:VIDeo:STANdard
Description
This command sets or queries the video standard.
Group
Trigger
Syntax 1
TRIGger:A:VIDeo:STANdard {CUStom|NTSc|PAL|
SCAM|HD480P60|HD720P60|HD1080I50|HD1080I60|
HD1080P24|HD1080P25|HD1080SF24}
Syntax 2
TRIGger:A:VIDeo:SOUrce?
Arguments
H CUStom
This argument sets the instrument to use custom video horizontal scan
rate parameters that you set with the TRIGger:A:VIDeo:SCAN com-
mand.
H NTSc
This argument sets the instrument to trigger on video signals that meet
the NTSC 525/60/2:1 standard (a line rate of 525 lines per frame and
a field rate of 60 Hz).
H PAL
This argument sets the instrument to trigger on video signals that meet
the NTSC 625/50/2:1 standard (a line rate of 625 lines per frame and
a field rate of 50 Hz).
H SECAM
This argument sets the instrument to trigger on video signals that meet
the SECAM standard.
H HD480P60
This argument sets the instrument to trigger on the HDTV 480/60
progressive format.
H HD720P60
This argument sets the instrument to trigger on the HDTV 720/60
progressive format.
H HD1080I50
This argument sets the instrument to trigger on HDTV 1080/50 inter-
laced format.
H HD1080I60

Table of Contents