EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #794 background imageLoading...
Page #794 background image
TDS5000B Online Programmer Guide
-- 7 7 4 --
TRIGger:A:VIDeo:STANdard
Description
This command sets or queries the video standard.
Group
Trigger
Syntax 1
TRIGger:A:VIDeo:STANdard {CUStom|NTSc|PAL|
SCAM|HD480P60|HD720P60|HD1080I50|HD1080I60|
HD1080P24|HD1080P25|HD1080SF24}
Syntax 2
TRIGger:A:VIDeo:SOUrce?
Arguments
H CUStom
This argument sets the instrument to use custom video horizontal scan
rate parameters that you set with the TRIGger:A:VIDeo:SCAN com-
mand.
H NTSc
This argument sets the instrument to trigger on video signals that meet
the NTSC 525/60/2:1 standard (a line rate of 525 lines per frame and
a field rate of 60 Hz).
H PAL
This argument sets the instrument to trigger on video signals that meet
the NTSC 625/50/2:1 standard (a line rate of 625 lines per frame and
a field rate of 50 Hz).
H SECAM
This argument sets the instrument to trigger on video signals that meet
the SECAM standard.
H HD480P60
This argument sets the instrument to trigger on the HDTV 480/60
progressive format.
H HD720P60
This argument sets the instrument to trigger on the HDTV 720/60
progressive format.
H HD1080I50
This argument sets the instrument to trigger on HDTV 1080/50 inter-
laced format.
H HD1080I60

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals