-- 6 1 --
*ESE
Description
This command sets and queries the bits in the Event Status Enable Reg-
ister (ESER). The ESER prevents events from being reported to the
Status Byte Register (STB). For a more detailed discussion of the use of
these registers, see Registers.
Group
Status and Error
Related Commands
*CLS, DESE, *ESR?, EVENT?, EVMsg?, *SRE, *STB?
Syntax 1
*ESE <NR1>
Syntax 2
*ESE?
Argument
H <NR1>
This specifies the binary bits of the ESER according to this value,
which ranges from 0 through 255.
The power--on default for the ESER is 0 if *PSC is 1. If *PSC is 0,
the ESER maintains the previous power cycle value through the cur-
rent power cycle.
Note: Setting the DESER and the ESER to the same values allows
only those codes to be entered into the Event Queue and summarized
on the ESB bit (bit 5) of the Status Byte Register. Use the DESE com-
mand to set the DESER.
Example 1
*ESE 209
This command sets the ESER to binary 11010001, which enables the
PON, URQ, EXE, and OPC bits.
Example 2
*ESE?
This query form of this command returns 186, showing that the ESER
contains the binary value 10111010.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals