TDS5000B Online Programmer Guide
-- 7 2 1 --
TRIGger:A:PULse:RUNT:THReshold:HIGH
Description
This command sets or queries the upper limit for the A pulse runt trig-
ger. This command is equivalent to selecting Runt Setup from the Trig
menu and setting the runt trigger Upper Level voltage.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:THReshold?
Syntax 1
TRIGger:A:PULse:RUNT:THReshold:HIGH <NR3>
Syntax 2
TRIGger:A:PULse:RUNT:THReshold:HIGH?
Argument
H <NR3>
This specifies the threshold value, in volts.
Example 1
TRIGger:A:PULse:RUNT:THReshold:HIGH?
This query might return
:TRIGGER:A:PULSE:RUNT:THRESHOLD:HIGH
1.1000E+00
indicating that the upper limit of the pulse runt trigger is currently set to
1.1 V.
Example 2
TRIGger:A:PULse:RUNT:THReshold:HIGH 120E-3
This command sets the upper limit of the pulse runt trigger to 120 mV.