EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 7 2 1 --
TRIGger:A:PULse:RUNT:THReshold:HIGH
Description
This command sets or queries the upper limit for the A pulse runt trig-
ger. This command is equivalent to selecting Runt Setup from the Trig
menu and setting the runt trigger Upper Level voltage.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:THReshold?
Syntax 1
TRIGger:A:PULse:RUNT:THReshold:HIGH <NR3>
Syntax 2
TRIGger:A:PULse:RUNT:THReshold:HIGH?
Argument
H <NR3>
This specifies the threshold value, in volts.
Example 1
TRIGger:A:PULse:RUNT:THReshold:HIGH?
This query might return
:TRIGGER:A:PULSE:RUNT:THRESHOLD:HIGH
1.1000E+00
indicating that the upper limit of the pulse runt trigger is currently set to
1.1 V.
Example 2
TRIGger:A:PULse:RUNT:THReshold:HIGH 120E-3
This command sets the upper limit of the pulse runt trigger to 120 mV.