TDS5000B Online Programmer Guide
-- 1 0 5 --
CALibrate:CALProbe:CH<x>?
Description
This query--only command instructs the instrument to perform a probe
calibration for the selected channel and returns the calibration status.
The Channel <x> range is 1 through 4. This command is equivalent to
selecting Probe Cal from the Vertical menu. You must warm up the in-
strument for at least 20 minutes before running this command.
Note: A probe query takes approximately 30 seconds to complete. No
other commands will be executed until calibration is complete.
Group
Calibration
Related Commands
CALibrate:PRObestate:CH<x>?, CH<x>:PROBECal?
Syntax
CALibrate:CALProbe:CH<x>?
Returns
H -1
Indicating the probe calibration failed for the selected channel.
H 0
Indicating the probe calibration passed for the selected channel.
H 1
Indicating the probe calibration is initialized for the selected channel.
Example
CALibrate:CALProbe:CH2?
This query might return
:CALIBRATE:CALPROBE:CH2 0
indicating the probe calibration passed for Channel 2.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals