-- 7 9 0 --
TRIGger:STATE?
Description
This query--only command returns the current state of the triggering
system. This command is equivalent to viewing the trigger status LEDs
on the instrument front panel.
Group
Trigger
Related Commands
TRIGger:A:MODe
Syntax
TRIGger:STATE?
Outputs
H ARMed
This indicates that the instrument is acquiring pretrigger information.
All triggers are ignored when TRIGger:STATE is arming.
H AUTO
This indicates that the instrument is in the automatic mode and ac-
quires data even in the absence of a trigger.
H DPO
This indicates that the instrument is in DPO mode.
H PARTial
This indicates that the A trigger has occurred and the instrument is
waiting for the B trigger to occur.
H REAdy
This indicates that all pretrigger information has been acquired and
that the instrument is ready to accept a trigger.
H SAVe
This indicates that the instrument is in save mode and is not acquiring
data.
H TRIGger
This indicates that the instrument triggered and is acquiring the post
trigger information.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals