EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 7 9 0 --
TRIGger:STATE?
Description
This query--only command returns the current state of the triggering
system. This command is equivalent to viewing the trigger status LEDs
on the instrument front panel.
Group
Trigger
Related Commands
TRIGger:A:MODe
Syntax
TRIGger:STATE?
Outputs
H ARMed
This indicates that the instrument is acquiring pretrigger information.
All triggers are ignored when TRIGger:STATE is arming.
H AUTO
This indicates that the instrument is in the automatic mode and ac-
quires data even in the absence of a trigger.
H DPO
This indicates that the instrument is in DPO mode.
H PARTial
This indicates that the A trigger has occurred and the instrument is
waiting for the B trigger to occur.
H REAdy
This indicates that all pretrigger information has been acquired and
that the instrument is ready to accept a trigger.
H SAVe
This indicates that the instrument is in save mode and is not acquiring
data.
H TRIGger
This indicates that the instrument triggered and is acquiring the post
trigger information.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Related product manuals