TDS5000B Online Programmer Guide
-- 11 0 --
CALibrate:RESults:SPC?
Description
This query--only command returns the results of the last SPC operation.
However, this query does not cause an SPC to be run. This command is
equivalent to selecting Instrument Calibration from the Utilities menu.
Group
Calibration
Related Commands
*CAL?
Syntax
CALibrate:RESults:SPC?
Example
CALibrate:RESults:SPC?
This query returns the results of the last SPC operation: either PASS or
FAIL.