-- 3 7 3 --
HORizontal:RECOrdlength
Description
This command sets the horizontal record length to the number of data
points in each frame. The query form of this command returns the cur-
rent horizontal record length. This command is equivalent to selecting
Resolution from the Horiz/Acq menu and then entering the desired Rec
Length.
Group
Horizontal
Syntax 1
HORizontal:RECOrdlength <NR1>
Syntax 2
HORizontal:RECOrdlength?
Argument
H <NR1>
This represents the supported values for horizontal record lengths,
which range from 500 through 400K data points. For more informa-
tion about valid data point ranges, select Specifications from the Help
menu and choose the Horizontal & Acquisition tab.
Example 1
HORizontal:RECOrdlength 50000
This command specifies that 50000 data points will be acquired for
each record.
Example 2
HORizontal:RECOrdlength?
This query might return:
:HORIZONTAL:RECOrdlength 5000
This indicates that the horizontal record length is equal to 5000 data
points.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals