TDS5000B Online Programmer Guide
-- 7 8 6 --
TRIGger:B:LEVel
Description
This command sets or queries the level for the B trigger. This command
is equivalent to selecting B Event (Delayed) Trigger Setup from the
Trig menu, selecting the A→B Seq tab and setting the B Trig Level
voltage.
Group
Trigger
Related Commands
TRIGger:A:LEVEL, TRIGger:B, TRIGger:B:EDGE:SOUrce
Syntax 1
TRIGger:B:LEVel {ECL|TTL|<NR3>}
Syntax 2
TRIGger:B:LEVel?
Arguments
H ECL
This specifies a preset ECL level of --1.3 V.
H TTL
This specifies a preset TTL level of 1.4 V.
H <NR3>
This is the B trigger level, in volts.
Example 1
TRIGger:B:LEVel?
This query might return
:TRIGGER:B:LEVEL 173.0000E-03
indicating that the B trigger level is currently set at 173 mV.
Example 2
TRIGger:B:LEVel ECL
This command sets the B trigger level to -- 1.3 V.