TRIGger:A:PULse:TIMEOut:TIMe
Description
This command sets or queries the pulse timeout trigger time (measured
in seconds). This command is equivalent to selecting Timeout Setup
from the Trig menu and setting a value for Timer.
Group
Trigger
Related Commands
TRIGger:A:PULse:TIMEOut?
Syntax 1
TRIGger:A:PULse:TIMEOut:TIMe <NR3>
Syntax 2
TRIGger:A:PULse:TIMEOut:TIMe?
Argument
H <NR3>
This argument specifies the timeout period in seconds.
Example 1
TRIGger:A:PULse:TIMEOut:TIMe?
This query might return
:TRIGGER:A:PULSE:TIMEOUT:TIME 2.0000E-9
indicating that the timeout is currently set to 2 ns.
Example 2
TRIGger:A:PULse:TIMEOut:TIMe 3.134E-6
This command sets the timeout to 3.134 µs.