-- 8 2 2 --
WFMOutpre:PT_Off?
Description
This query--only command returns the trigger point relative to
DATa:STARt for the waveform specified by the DATa:SOUrce com-
mand.
Note: This returned value is the point immediately following the actu-
al trigger.
Group
Waveform Transfer
Related Commands
DATa:SOUrce, DATa:STARt, WFMOutpre:XZEro
Syntax
WFMOutpre:PT_Off?
Example
WFMOutpre:PT_Off?
This query might return
:WFMOUTPRE:PT_OFF 251
specifying that the trigger actually occurred between points 250 and
251.